performance feature

description

flow rate (measuring point tube attachment)
max. 8 l/min (low flow)
vacuum
-60 mbar to -350 mbar (in steps of 5 mbar) (medium vacuum)
conversion factor: 10 mbar ~ 1 kPa ~ 7.5 mmHg
canister
disposable secretion canister (250 ml)
suction tube
Ø 4 mm (internal), length 150 cm, incl. step connector
power supply unit
ATM024T-W120V, cable length 4 m
nominal voltage of the power supply unit
In: AC 100 – 240 V~ / 50-60 Hz / 580-320 mA
Out: DC 12 V / 2.0 A
maximum load current
2.0 A
nominal voltage of printed circuit board
12 V
power consumption at 12 V
24 W
current consumption at 12 V
2.0 A
degree of protection pursuant to IEC 60601-1
Type BF
risk classification pursuant to 93/42/EEC, Annex IX
IIa
protection class pursuant to IEC 60601-1
II
IP degree of protection
IP33
CE marking
CE0494
sound emission
operation: 35 dB(A)
high priority alarm: 53 dB(A)
low priority alarm: 52 dB(A)
ambient conditions
transport/storage: -25°C to +60°C, humidity of max. 93% non-condensing
operation: +5°C to +40°C, humidity 15% to 93% non-condensing
air pressure: 700 hPa to 1060 hPa
battery, rechargeable
7.4 V; 4.4 Ah (lithium ion battery)
charging time if battery is empty
6 – 7 hours
energy of battery pack
<80 Wh
dimensions (HxWxD) [in mm]
165 x 220 x 90
weight (basic unit)
1.2 kg
pressure display accuracy
target pressure > -120 mbar: max. 5% deviation
target pressure ≤ -120 mbar: max. 10% deviation
operating time
continuous operation
runtime in battery operation
approx. 18 hours, depending on the strain of the motor
item number
100567-3-EW